description2
SSY-ED Pure Water Treatment System
Product Introduction
Hospital use pure water treatment system as water for the cleaning and disinfection equipment of the endoscopic center, the conductivity is ≤15, the recovery rate is 70%, the desalination rate is 99%, the filter membrane aperture used in the production of purified water is ≤0.2, and the bacteria is < 10CFU/100ml, which meets the pure water quality standard of the national “WS507-2016” technical specification. Issue water quality monitoring report.
According to standard “WS507-2016”.
Application: Endoscopic cleaning pure water equipment is suitable for Xinhua, Laoken, Jianghan, Kunshan, Qianying, White elephant, Meimei, Kengwang and other hospitals endoscopic cleaning equipment supporting pure water.
Technical Features
1. With patented circulating disinfection and sterilization process, the water quality meets the latest requirements of endoscopic cleaning water.
2. PLC intelligent man-machine control system,convenient operation, real-time monitoring equipment operation information.
3. Automatic disinfection mode,easy to operate with one click.
4. Multiple detection instrument,Real-time online monitoring of water quality, pressure and flow information.
5. Regular cycle flushing discharge, effectively avoid the formation of stagnant water.
6. Automatic water leakage detection sensor, water leakage in time to cut off water.
7. Complete historical data recording function, automatically save water quality data, support query and download.
Model |
Capacity |
Power |
Input Power |
Overall Size |
SSY-ED-200L |
200L/H |
2.2KW |
AC220V/50HZ |
1100*730*1600mm |
SSY-ED-250L |
250L/H |
2.2KW |
AC220V/50HZ |
1100*730*1600mm |
SSY-ED-300L |
300L/H |
2.2KW |
AC220V/50HZ |
1100*730*1600mm |
SSY-ED-500L |
500L/H |
2.5KW |
AC220V/50HZ |
1100*730*1600mm |
Technical Parameter
Model: |
SSY-ED-200/250/300/500L/H |
Capacity: |
200/250/300/500L/H |
Environment Temperature: |
5-45℃ |
Raw Water Pressure: |
0.2-0.4Mpa |
Productive Water Quality Index: |
RO Conductivity<15μs/cm@25℃;Total Bacteria ≤10cfu/100ml Bacterial Removal Rate≥99.6% Particle (>0.2μm) <1 piece/ml Ion Retention Rate:97%-99% Organic matter retention rate>99% It meets the technical specification of WS507-2016 soft endoscope cleaning and disinfection. |
Application: |
hospitals endoscopic cleaning equipment |
SSY-ED Process Flow Diagram
Raw water → Raw water pump → Quartz sand tank → Activated carbon tank → softening tank → Salt tank → Precision filter → High pressure pump → Reverse osmosis membrane → Pure water tank→ Delivery pump → Water point