description2
SSY-UP2.0 Laboratory Ultra Pure Water System
Product Introduction
The desktop laboratory ultra pure water system adopts reverse osmosis technology and ion exchange technology combined with microcomputer single-board computer program control, water quality detection automatic display, so as to obtain high-quality produced water, its effluent resistivity can generally reach 18MΩ/cm. Two-stage pretreatment design, one-stage filtration is quartz sand to remove solid pollutants such as rust, sediment and suspended matter. Secondary filtration pp cotton removes residual chlorine, heterochromatic odor, organic matter and other pollutants in water.
Better than water standards “GB/T6682-2008”,“GB/T33087-2016”,”“GB/T11446.1-2013”, “ISO3696, ASTM, D1193, D1125-95(1995), CLSI”, “EP,CAP, CHP, USP”and so on.
Application: Daily analytical water,Preparation solution flushing,Preparation reagent,Utensil washing.
Technical Features
1. Dual water intake mode,at the same time, take type I UP ultra-pure water and type III RO pure water.
2. Perfect online monitoring and alarm system, real-time display of water quality.
3. The use of NSF certified pipes and joints greatly reduces the amount of TOC precipitation in the system and ensures the quality of produced water.
4. Using imported double wavelength ultraviolet sterilizer, efficient sterilization, reduce TOC content.
5. PES polysulfone composite membrane filter element is imported to ensure the end effluent quality asepsis.
Model |
Capacity |
Power |
Input Power |
Overall Size |
SSY-UP2.0-10L/20L |
10/20L/H |
100W |
AC220V/50HZ |
520*355*590mm |
Technical Parameter
Model: |
SSY-UP2.0-10L/20L |
Capacity: |
10/20L/H 25℃ (For every 1℃drop in temperature, water production decreases by about 3%.) |
Resistivity 25℃: |
15-18.2MΩ.cm |
TOC: |
<5ppb |
Heavy metal ion: |
<0.1ppb |
Bacteria: |
<1 CFU/ml |
Particulate matter(>0.1μm): |
<1 piece/ml |
Ion Retention Rate: |
97-99% |
Organic matter retention rate: |
>99%,when MW>2000Dalton |
Particle and bacterial retention rate: |
>99% |
Raw water requirement: |
Meet GB5749-2006 urban drinking tap water, water temperature 5-45°C, water pressure 0.2-0.4Mpa |
Size: |
520*355*590mm |
Electrical requirement: |
AC220V/50HZ |
Power: |
100W |
Application: |
Daily analytical water,Preparation solution flushing,Preparation reagent,Utensil washing. |
SSY-UP2.0 Process Flow Diagram
Raw water →CTO filter →PP filter → booster pump → reverse osmosis membrane → Purification system →UV sterilization →UF ultrafiltration system →PES terminal microfiltration → water point